| Item Number |
CLS-1191-SQL-1010C |
| System |
Automatic high-pressure steam sterilization |
| Operating Temperature Range |
45~135°C |
| |
45 ~80°C (pre-heating) / 45 ~ 60°C (heat retention) / 65 ~100°C(melting) / 105~135°C (sterilization |
| Maximum Working Pressure |
0.255MPa |
| Lid (cover mechanism) |
Manual upward opening with an interlock for safety |
| Heater |
2000W x 2 units |
| Exhaust Valve |
One exhaust valve and one slow-release valve |
| Option port |
For sample sensor (1/4), recorder (1/4) and connection to pressure gauge (branched from the electromagnetic exhaust duct) |
| Cooling fan |
Axial fan motor |
| Temperature Controller |
PID control by microprocessor |
| Temperature Display/Setting |
Digital display / digital setting by ▲▼ keys |
| Timer/Timer Resolution |
0 or 1 min. to 99 hrs 59 min. / 1 minute |
| Operation Mode |
Instrument sterilization course, liquid sterilization course, sterilization and keep warm course, melting and keep warm course, customer-programmed course |
| Other features |
Key locking, presetting, saving, preheating, forced cooling, sample temperature sensor (option), pattern locking, up to 20 error log saving, display of accumulated working time / present time, ON-OFF beeping setting |
| Safety device |
Sensor failure detection, SSR short-circuit, broken heater wire, prevention of idle heating (liquid expansion type), alarm against the absence of a drain bottle, failure in locking the lid, memory error detection, pressure relief valve (0.255 MPa) |
| External dimensions |
680 x 760 x 1154 mm WxDxH |
| Internal dimensions of chamber |
Ø450 x D692 mm |
| Internal capacity |
110L |
| Weight |
~170kg |
| Power source |
AC220V (19A) (no plug, round terminal) |
| Included accessories |
2 pcs. stainless steel mesh basket (ø424xH300mm) |
| |
Vapor cup x 1, Drain bottle x 1, Drain board x 1, Filter x 1 |